Your Cart
(empty)
0
Tradeport Electronics
measurement is our business
Cart : (empty)

Soken DAC-PBC-8 - Electronic Burden for Current Transformer

Electronic Burden for Current Transformer

Soken

Category : Instrument Transformer Testing Set

by quote only Add to Cart

Documents:
Brochure

Overview
Features
Information
Calibration
  • Programmable burden used for current error testing in accordance with international standard IEC 60044-1.
  • Testing current can be varied over a wide range: from 1 % to 200% of the rated secondary current.
  • Rated burdens from 0 to 50VA are programmable.
  • USB Interface is a standard fixture.
  • Lead wire compensation (the maximum of 0.2 Ω) by connecting with four terminals is possible.
  • Max real burden is up to 400VA with option units. (when 200% at rated secondary current 1A)

 

4-terminal Connecting diagrams

 

Option Electric Burden

Optional burden DAC-FBC series are available in need of further VA.

  • DAC-FBC Series: : 50VA, 100VA and 200VA type

Combining DAC-PBC-8 with all FBC series, max 400VA* is available. *(when 200% at rated secondary current 1A)

 

Instrument Transformer Test Set DAC-VCTT-8

Combining with DAC-VCTT-8, Current Burden DAC-PBC-8 can be controlled from DAC-VCTT-8. Click here to find out more about the DAC-VCTT-8.

Soken DAC-PBC-8 - Electronic Burden for Current Transformer
  • Programmable burden used for current error testing in accordance with international standard IEC 60044-1.
  • Testing current can be varied over a wide range: from 1 % to 200% of the rated secondary current.
  • Rated burdens from 0 to 50VA are programmable.
  • USB Interface is a standard fixture.
  • Lead wire compensation (the maximum of 0.2 Ω) by connecting with four terminals is possible.
  • Max real burden is up to 400VA with option units. (when 200% at rated secondary current 1A)

 

General

Rated Secondary Current1A and 5A
Test Current range1% to 200% of the rated secondary current
Rated burden setting 0.000 to 50.0 VA
Maximum of 400VA* with optional Electric burden *1% to 200% of the rated secondary current
Burden Power factorDelay 0.5 to 1
Lead wire compensation0.000 to 0.200Ω
Test Frequency50/60 Hz
Settable resolution
Rated burden setting0 – 4 VA: 0.001 VA steps
4 – 20 VA: 0.01 VA steps
20 – 50 VA: 0.1 VA steps
Burden Power Factor0.01 steps
Accuracy (target value)Burden and Burden Power Factor: ±4%
CommunicationUSB (2.0/1.1)
Other Functions Test condition programming, over current protection, Alarm function for measuring circuit, lead wire compensation
Rated Supply Voltage100 to 240V ±10%
Dimensions W430 x H250 x D380mm (without projections)

Do you have a Electronic Burden for Current Transformer that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Electronic Burden for Current Transformer calibrated today! Click Here!