GW Instek DAQ-9600G - Data Acquisition System (Mainframe) with GPIB

Data Acquisition System (Mainframe) with GPIB

GW Instek

Rated 0 / 5 (0 reviews)

Category : Data Acquisition

by quote only add to cart

Documents:
Datasheet Modules Datasheet

Overview
Features
Specifications
Calibration

The DAQ-9600 data acquisition system is a modularized data acquisition system with high flexibility and higher performance. The mainframe is equipped with 3 module slots and a built-in precision 6 1/2 digital DMM is the core of its test and measurement. 5 modules are available to meet different measurement needs. For the research and development of analyzing product characteristics or the production and manufacturing of system testing or fault diagnosis, a data acquisition system with flexibility and high performance can effectively fulfill different measurement requirements by expansion and change that make the overall test simpler, faster and more reliable.

 

Customized configuration for changing needs

The DAQ-9600 provides 3 module slots, and all of which can be used to install and control selected modules. 5 different modules are available to meet different measurement needs. These modules include general-purpose modules, multi-function modules, matrix module and high-voltage and high-current module. Users can choose, mix and match these modules to obtain the functions as required. The measurement channels of each module can be configured independently, including measurement function, range, speed and the advanced measurement functions such as offset compensation. The DAQ-9600 can provide up to 60 two-wire measurements, 120 single-ended measurements, or 96 matrix crosspoints in a half-rack, which saves work space and cost when completing complex and diverse measurement tasks.

 

p1-1.png

 

Precise test and measurement

The test and measurement core of the DAQ-9600 is a 6 1/2-digit DMM, which has the functions and performance of an industrial-grade DMM. It can measure and convert 14 different input signals. In addition to high resolution, it also provides basic accuracy of 0.0035% DC voltage, 0.05% basic accuracy of AC voltage and 0.01% basic accuracy of resistance (within one-year calibration period). The functions of this DMM can be used with general-purpose modules (DAQ-900/ DAQ-901), multi-function module (DAQ-903) and high-voltage and high-current module (DAQ-909). All multiplexer modules use "disconnect before connect" scanning to ensure that only one channel (or a pair of channels) is closed at a time.

 

 

p2.png

 

 

Built-in signal conditioning

The DAQ -9600 does not rely on external or plug-in signal conditioning modules to handle functions other than DC voltage, which not only greatly reduces noise and errors generated by external wiring from entering the system, and improves reliability. At the same time, it does not have unnecessary cables, wire distribution boxes and signal conditioning components that makes configuration quicker and easier. Taking temperature measurement as an example, the measurement of thermocouples is often accompanied by the Seebeck effect (the thermoelectric phenomenon in which the temperature difference between two different electrical conductors or semiconductors causes a voltage difference between two substances). All DAQ-9600 modules provide cold junction compensation (except DAQ-904) to ensure maximum accuracy for all channels measuring thermocouples, avoiding the use of temperature extension wires with external cold junctions, which will increase noise and errors generated by external wiring into the system.

 

Internal memory and USB storage

The DAQ-9600 can store the measurement results without connecting to a PC with the non-volatile memory capacity of 100k readings. Users can store the measurement results inside the system first, and then connect to a PC to collect data. In addition, for larger memory capacity, the DAQ-9600 also supports the USB storage function, which can directly record data to an external USB flash drive, or copy the data of the internal non-volatile memory to the flash drive for follow-up analysis.

 

Simplify data collection and analysis

The DAQ-9600 provides LAN, USB and GPIB (optional) to meet the most suitable and customary remote control interface requirements. Users can be benefitted from the free data collection software (DAQ-Data Logger) to easily and quickly set up tests, acquire and store measurement data, and perform real-time observation and analysis during the measurement process. If LAN is used as the interface, it can be easily connected to the company network to collect measurement data and store it in the central database. It can also use a common web browser (such as Chrome, IE) through the built-in graphical web interface to configure test and measurements, define and execute scan lists to reduce the time spent on programming.

 

  • Large 4.3”TFT color display
  • 3-slot mainframes with built-in 6 ½ digit DMM
  • Basic 0.0035% DCV accuracy
  • 5 selectable switch modules
  • Up to DC 600V / AC 400V voltage measurement (DAQ-909 module)
  • Up to 450 channel/s scan rate
  • Up to 100 kilo points internal memory
  • Measures and converts 14 different input signals:
  • Temperature with thermocouple, RTDs and thermistor; dc/ac volts; 2- and 4-wire resistance; frequency and period; dc/ac current and capacitance; direct strain and bridge strain
  • USB storage support to copy/log data in standalone operation
  • Interface: Digit I/O, LAN, USB host/device and mini GPIB (optional)

 

DC Characteristics - Accuracy : ± (% of reading + % of range)

DC Voltage (1)
Range 100.0000 mV (2)
Resolution0.1μV
Input Resistance etc.10MΩ or >10GΩ
24 Hour TCAL± 1°C0.0030 + 0.0050
90 Day TCAL± 5°C0.0040 + 0.0060
1 Year TCAL± 5°C0.0050 + 0.0060
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0005 + 0.0005
Range 1.000000 V (2)
Resolution1μV
Input Resistance etc.10MΩ or >10GΩ
24 Hour TCAL± 1°C0.0020 + 0.0006
90 Day TCAL± 5°C0.0035 + 0.0007
1 Year TCAL± 5°C0.0048 + 0.0007
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0005 + 0.0001
Range 10.00000 V (2)
Resolution10μV
Input Resistance etc.10MΩ or >10GΩ
24 Hour TCAL± 1°C0.0015 + 0.0004
90 Day TCAL± 5°C0.0020 + 0.0005
1 Year TCAL± 5°C0.0035 + 0.0005
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0005 + 0.0001
Range 100.0000 V (2)
Resolution0.1mV
Input Resistance etc.10MΩ±1%
24 Hour TCAL± 1°C0.0020 + 0.0006
90 Day TCAL± 5°C0.0035 + 0.0006
1 Year TCAL± 5°C0.0050 + 0.0006
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0005 + 0.0001
Range 600.000 V (2)
Resolution1mV
Input Resistance etc.10MΩ±1%
24 Hour TCAL± 1°C0.0025 + 0.0020
90 Day TCAL± 5°C0.0040 + 0.0020
1 Year TCAL± 5°C0.0050 + 0.0020
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0005 + 0.0001
Resistance (1)(3)
Range 100.0000 Ω (2)
Resolution100μΩ
Input Resistance etc.1mA
24 Hour TCAL± 1°C0.003 + 0.0030
90 Day TCAL± 5°C0.008 + 0.004
1 Year TCAL± 5°C0.010 + 0.004
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0008 + 0.0005
Range 1.000000 kΩ (2)
Resolution1mΩ
Input Resistance etc.1mA
24 Hour TCAL± 1°C0.002 + 0.0005
90 Day TCAL± 5°C0.008 + 0.001
1 Year TCAL± 5°C0.010 + 0.001
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0008 + 0.0001
Range 10.00000 kΩ (2)
Resolution10mΩ
Input Resistance etc.100μA
24 Hour TCAL± 1°C0.002 + 0.0005
90 Day TCAL± 5°C0.008 + 0.001
1 Year TCAL± 5°C0.010 + 0.001
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0008 + 0.0001
Range 100.0000 kΩ (2)
Resolution100mΩ
Input Resistance etc.10μA
24 Hour TCAL± 1°C0.002 + 0.0005
90 Day TCAL± 5°C0.008 + 0.001
1 Year TCAL± 5°C0.010 + 0.001
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0008 + 0.0001
Range 1.000000 MΩ (2)
Resolution
Input Resistance etc5μA
24 Hour TCAL± 1°C0.002 + 0.0010
90 Day TCAL± 5°C0.008 + 0.001
1 Year TCAL± 5°C0.010 + 0.001
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0010 + 0.0002
Range 10.00000 MΩ (2)
Resolution10Ω
Input Resistance etc.100μA
24 Hour TCAL± 1°C0.002 + 0.0005
90 Day TCAL± 5°C0.008 + 0.001
1 Year TCAL± 5°C0.010 + 0.001
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.0008 + 0.0001
Range 100.0000 MΩ (2)
Resolution100Ω
Input Resistance etc.500nA//10MΩ
24 Hour TCAL± 1°C0.300 + 0.0100
90 Day TCAL± 5°C0.800 + 0.010
1 Year TCAL± 5°C0.800 + 0.010
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.1500 + 0.0004
Range 1.000000 GΩ (2)
Resolution1kΩ
Input Resistance etc.500nA//10MΩ
24 Hour TCAL± 1°C2.500 + 0.0500
90 Day TCAL± 5°C3.500 + 0.050
1 Year TCAL± 5°C3.500 + 0.050
Temperature Coefficient 0°~ 18°C / 28°~ 55°C1.0000 + 0.0040
DC Current (1)
Range 1.000000 μA (2)
Resolution1pA
Input Resistance etc.< 0.015 V
24 Hour TCAL± 1°C0.025 + 0.050
90 Day TCAL± 5°C0.050 + 0.050
1 Year TCAL± 5°C0.050 + 0.050
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.003
Range 10.00000 μA (2)
Resolution10pA
Input Resistance etc.< 0.15 V
24 Hour TCAL± 1°C0.020 + 0.010
90 Day TCAL± 5°C0.040 + 0.025
1 Year TCAL± 5°C0.050 + 0.025
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.003
Range 100.0000 μA (2)
Resolution100pA
Input Resistance etc.< 0.020 V
24 Hour TCAL± 1°C0.010 + 0.020
90 Day TCAL± 5°C0.040 + 0.025
1 Year TCAL± 5°C0.050 + 0.025
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.003
Range 1.000000 mA (2)
Resolution1nA
Input Resistance etc.< 0.20 V
24 Hour TCAL± 1°C0.007 + 0.006
90 Day TCAL± 5°C0.030 + 0.006
1 Year TCAL± 5°C0.050 + 0.006
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.001
Range 10.00000 mA (2)
Resolution10nA
Input Resistance etc.< 0.15 V
24 Hour TCAL± 1°C0.007 + 0.020
90 Day TCAL± 5°C0.030 + 0.020
1 Year TCAL± 5°C0.050 + 0.020
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.002
Range 100.0000 mA (2)
Resolution100nA
Input Resistance etc.< 0.7 V
24 Hour TCAL± 1°C0.010 + 0.004
90 Day TCAL± 5°C0.030 + 0.005
1 Year TCAL± 5°C0.050 + 0.005
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.002 + 0.001
Range 2.000000 A (2)
Resolution1μA
Input Resistance etc.< 0.8 V
24 Hour TCAL± 1°C0.180 + 0.020
90 Day TCAL± 5°C0.200 + 0.020
1 Year TCAL± 5°C0.200 + 0.020
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 + 0.001
Diode Test (1)(4)
Range (2) 5.00000 V
Resolution10μV
Input Resistance etc.1 mA
24 Hour TCAL± 1°C0.002 + 0.030
90 Day TCAL± 5°C0.008 + 0.030
1 Year TCAL± 5°C0.01 + 0.030
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.001 + 0.002

 

AC Characteristics - Accuracy : ± (% of reading + % of range)

True RMS AC Voltage (5)(6)(7)(8)
Range 100.0000 mV (2)
Resolution 0.1μV
Input Resistance etc. (3Hz - 5Hz)
24 Hour TCAL± 1°C1.00 + 0.03
90 Day TCAL± 5°C1.00 + 0.04
1 Year TCAL± 5°C1.00 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.004
Input Resistance etc. (5Hz - 10Hz)
24 Hour TCAL± 1°C0.35 + 0.03
90 Day TCAL± 5°C0.35 + 0.04
1 Year TCAL± 5°C0.35 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.004
Input Resistance etc. (10Hz - 20kHz)
24 Hour TCAL± 1°C0.04 + 0.03
90 Day TCAL± 5°C0.05 + 0.04
1 Year TCAL± 5°C0.06 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 + 0.003
Input Resistance etc. (20kHz - 50kHz)
24 Hour TCAL± 1°C0.10 + 0.05
90 Day TCAL± 5°C0.11 + 0.05
1 Year TCAL± 5°C0.12 + 0.05
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.011 + 0.005
Input Resistance etc. (50kHz - 100kHz)
24 Hour TCAL± 1°C0.55 + 0.08
90 Day TCAL± 5°C0.60 + 0.08
1 Year TCAL± 5°C0.60 + 0.08
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.060 + 0.008
Input Resistance etc. (100kHz -300kHz)
24 Hour TCAL± 1°C4.00 + 0.50
90 Day TCAL± 5°C4.00 + 0.50
1 Year TCAL± 5°C4.00 + 0.50
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.200 + 0.020
Range 1.000000 V to 400.000 V (2)
Resolution 1μV ~ 1mV
Input Resistance etc. (3Hz - 5Hz)
24 Hour TCAL± 1°C1.00 + 0.02
90 Day TCAL± 5°C1.00 + 0.03
1 Year TCAL± 5°C1.00 + 0.03
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.004
Input Resistance etc. (5Hz - 10Hz)
24 Hour TCAL± 1°C0.35 + 0.02
90 Day TCAL± 5°C0.35 + 0.03
1 Year TCAL± 5°C0.35 + 0.03
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.004
Input Resistance etc. (10Hz - 20kHz)
24 Hour TCAL± 1°C0.04 + 0.02
90 Day TCAL± 5°C0.05 + 0.03
1 Year TCAL± 5°C0.06 + 0.03
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 + 0.003
Input Resistance etc. (20kHz - 50kHz)
24 Hour TCAL± 1°C0.10 + 0.04
90 Day TCAL± 5°C0.11 + 0.05
1 Year TCAL± 5°C0.12 + 0.05
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.011 + 0.005
Input Resistance etc. (50kHz - 100kHz)
24 Hour TCAL± 1°C0.55 + 0.08
90 Day TCAL± 5°C0.60 + 0.08
1 Year TCAL± 5°C0.60 + 0.08
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.060 + 0.008
Input Resistance etc. (100kHz -300kHz)
24 Hour TCAL± 1°C4.00 + 0.50
90 Day TCAL± 5°C4.00 + 0.50
1 Year TCAL± 5°C4.00 + 0.50
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.200 + 0.020
True RMS AC Current (5)(7)(9)
Range 100.0000 μA (2)
Resolution < 0.020 V
Input Resistance etc. (3Hz – 5Hz)
24 Hour TCAL± 1°C1.00 + 0.04
90 Day TCAL± 5°C1.00 + 0.06
1 Year TCAL± 5°C1.00 + 0.06
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.006
Input Resistance etc. (5Hz – 10Hz)
24 Hour TCAL± 1°C0.35 + 0.04
90 Day TCAL± 5°C0.35 + 0.06
1 Year TCAL± 5°C6 0.35 + 0.06
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Input Resistance etc. (10Hz – 5kHz)
24 Hour TCAL± 1°C0.10 + 0.04
90 Day TCAL± 5°C0.10 + 0.06
1 Year TCAL± 5°C0.10 + 0.06
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015 + 0.006
Input Resistance etc. (5kHz – 10kHz)
24 Hour TCAL± 1°C0.18 + 0.04
90 Day TCAL± 5°C0.18 + 0.10
1 Year TCAL± 5°C0.18 + 0.10
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Range 1.000000 mA (2)
Resolution < 0.20 V
Input Resistance etc. (3Hz – 5Hz)
24 Hour TCAL± 1°C1.00 + 0.04
90 Day TCAL± 5°C1.00 + 0.04
1 Year TCAL± 5°C1.00 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.006
Input Resistance etc. (5Hz – 10Hz)
24 Hour TCAL± 1°C0.30 + 0.04
90 Day TCAL± 5°C0.30 + 0.04
1 Year TCAL± 5°C0.30 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Input Resistance etc. (10Hz – 5kHz)
24 Hour TCAL± 1°C0.10 + 0.04
90 Day TCAL± 5°C0.10 + 0.04
1 Year TCAL± 5°C0.10 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015 + 0.006
Input Resistance etc. (5kHz – 10kHz)
24 Hour TCAL± 1°C0.15 + 0.04
90 Day TCAL± 5°C0.15 + 0.04
1 Year TCAL± 5°C0.15 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.030 + 0.006
Range 10.00000 mA (2)
Resolution < 0.15 V
Input Resistance etc. (3Hz – 5Hz)
24 Hour TCAL± 1°C1.00 + 0.04
90 Day TCAL± 5°C1.00 + 0.04
1 Year TCAL± 5°C1.00 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.006
Input Resistance etc. (5Hz – 10Hz)
24 Hour TCAL± 1°C0.35 + 0.04
90 Day TCAL± 5°C0.35 + 0.04
1 Year TCAL± 5°C0.35 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Input Resistance etc. (10Hz – 5kHz)
24 Hour TCAL± 1°C0.10 + 0.04
90 Day TCAL± 5°C0.10 + 0.04
1 Year TCAL± 5°C0.10 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015 + 0.006
Input Resistance etc. (5kHz – 10kHz)
24 Hour TCAL± 1°C0.18 + 0.04
90 Day TCAL± 5°C0.18 + 0.04
1 Year TCAL± 5°C0.18 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.030 + 0.006
Range 100.0000 mA (2)
Resolution < 0.7 V
Input Resistance etc. (3Hz – 5Hz)
24 Hour TCAL± 1°C1.00 + 0.04
90 Day TCAL± 5°C1.00 + 0.04
1 Year TCAL± 5°C1.00 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.006
Input Resistance etc. (5Hz – 10Hz)
24 Hour TCAL± 1°C0.30 + 0.04
90 Day TCAL± 5°C0.30 + 0.04
1 Year TCAL± 5°C0.30 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Input Resistance etc. (10Hz – 5kHz)
24 Hour TCAL± 1°C0.10 + 0.04
90 Day TCAL± 5°C0.10 + 0.04
1 Year TCAL± 5°C0.10 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015 + 0.006
Input Resistance etc. (5kHz – 10kHz)
24 Hour TCAL± 1°C0.15 + 0.04
90 Day TCAL± 5°C0.15 + 0.04
1 Year TCAL± 5°C0.15 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.030 + 0.006
Range 2.000000 A (2)
Resolution < 0.8 V
Input Resistance etc. (3Hz – 5Hz)
24 Hour TCAL± 1°C1.00 + 0.04
90 Day TCAL± 5°C1.00 + 0.04
1 Year TCAL± 5°C1.00 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100 + 0.006
Input Resistance etc. (5Hz – 10Hz)
24 Hour TCAL± 1°C0.35 + 0.04
90 Day TCAL± 5°C0.35 + 0.04
1 Year TCAL± 5°C0.35 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035 + 0.006
Input Resistance etc. (10Hz – 5kHz)
24 Hour TCAL± 1°C0.23 + 0.04
90 Day TCAL± 5°C0.23 + 0.04
1 Year TCAL± 5°C0.23 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015 + 0.006
Input Resistance etc. (5kHz – 10kHz)
24 Hour TCAL± 1°C0.23 + 0.04
90 Day TCAL± 5°C0.23 + 0.04
1 Year TCAL± 5°C0.23 + 0.04
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.030 + 0.006

 

Frequency and Period Characteristics - Accuracy : ± (% of reading)

Frequency / Period (9)(10)(11)(12)
Range 100.0000mV to 400.000V (2)
Input Resistance etc. (3Hz - 5Hz)
24 Hour TCAL± 1°C0.050
90 Day TCAL± 5°C0.050
1 Year TCAL± 5°C0.050
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035
Input Resistance etc. (5Hz - 10Hz)
24 Hour TCAL± 1°C0.050
90 Day TCAL± 5°C0.050
1 Year TCAL± 5°C0.050
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.035
Input Resistance etc. (10Hz - 40Hz)
24 Hour TCAL± 1°C0.030
90 Day TCAL± 5°C0.030
1 Year TCAL± 5°C0.030
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015
Input Resistance etc. (40Hz -1MHz)
24 Hour TCAL± 1°C0.006
90 Day TCAL± 5°C0.006
1 Year TCAL± 5°C0.006
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.015
24 Hour TCAL± 1°C0.100
90 Day TCAL± 5°C0.100
1 Year TCAL± 5°C0.100
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.100

 

Temperature Characteristics

Temperature (RTD) (13)
Range -200 °C ~ -100 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.09 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.004 °C / °C
Range -100 °C ~ -20 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.08 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 °C / °C
Range -20 °C ~ 20 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.06 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 °C / °C
Range 20 °C ~ 100 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.08 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.005 °C / °C
Range 100 °C ~ 300 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.12 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.007 °C / °C
Range 300 °C ~ 600 °C (2)
Resolution0.001 °C
1 Year TCAL± 5°C0.22 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.009 °C / °C
Temperature (Thermocouples) (13)
Range -200 to +1000 °C (2)
Resolution0.002 °C
Input Resistance etc.E
1 Year TCAL± 5°C0.2 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.03 °C / °C
Range -210 to +1200 ˚C (2)
Resolution0.002 °C
Input Resistance etc.J
1 Year TCAL± 5°C0.2 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.03 °C / °C
Range -200 to +400 °C (2)
Resolution0.002 °C
Input Resistance etc.T
1 Year TCAL± 5°C0.3 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.04 °C / °C
Range -200 to +1372 ˚C (2)
Resolution0.002 °C
Input Resistance etc.K
1 Year TCAL± 5°C0.3 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.04 °C / °C
Range -200 to +1300 ˚C (2)
Resolution0.003 °C
Input Resistance etc.N
1 Year TCAL± 5°C0.4 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 °C / °C
Range -50 to +1768 ˚C (2)
Resolution0.01 °C
Input Resistance etc.R
1 Year TCAL± 5°C1 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.14 °C / °C
Range -50 to +1768 ˚C (2).
Resolution0.01 °C
Input Resistance etc.S
1 Year TCAL± 5°C1 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.14 °C / °C
Range +350 to +1820 ˚C (2)
Resolution0.01 °C
Input Resistance etc.B
1 Year TCAL± 5°C1 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.14 °C / °C
Temperature (Thermistor) (13)
Range –80 ° to 150 °C (2)
Resolution 0.01 °C
1 Year TCAL± 5°C0.01 °C
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.003 °C / °C

 

Capacitance Characteristics - Accuracy : ± (% of reading + % of range)

Capacitance (14)
Range 1.000 nF (2)
24 Hour TCAL± 1°C2.00 + 2.00
90 Day TCAL± 5°C2.00 + 2.00
1 Year TCAL± 5°C2.00 + 2.00
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01
Range 10.00 nF (2)
24 Hour TCAL± 1°C2.00 + 1.00
90 Day TCAL± 5°C2.00 + 1.00
1 Year TCAL± 5°C2.00 + 1.00
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01
Range 100.0 nF (2)
24 Hour TCAL± 1°C2.00 + 0.40
90 Day TCAL± 5°C2.00 + 0.40
1 Year TCAL± 5°C2.00 + 0.40
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01
Range 1.000 μF (2)
24 Hour TCAL± 1°C2.00 + 0.40
90 Day TCAL± 5°C2.00 + 0.40
1 Year TCAL± 5°C2.00 + 0.40
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01
Range 10.00 μF (2)
24 Hour TCAL± 1°C2.00 + 0.40
90 Day TCAL± 5°C2.00 + 0.40
1 Year TCAL± 5°C2.00 + 0.40
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01
Range 100.0 μF (2)
24 Hour TCAL± 1°C2.00 + 0.40
90 Day TCAL± 5°C2.00 + 0.40
1 Year TCAL± 5°C2.00 + 0.40
Temperature Coefficient 0°~ 18°C / 28°~ 55°C0.05 + 0.01

 

General

Display 4.3" color WQVGA (480x272) with LED backlight
Interface RS -232C, USB host/device, LAN, Digital I/O; GPIB(optional)
Power Source AC 100 V / 120 V / 220 V / 240 V ±10%
Power Line Frequency50 Hz / 60 Hz ±10%
Power Consumption Max. 50VA
Dimensions220(W) x 88(H) x 348.6(D) mm ~ without bumper
266.9(W) x 107(H) x 357.8(D) mm ~ with bumper
WeightApprox. 4.5kg

[1]. DC Specification: In addition to the availability that requires warm-up of 60 minutes, it must be set in 5/s speed rate, AZero on.
[2]. The entire range of measurement will pass the set range by 20% except the tests of 600 V DC, 400 V AC, 2 A DC, 2 A AC and diode.
[3]. This specifications applies to 4-wire ohms function or 2-wire ohms using math null for offset. Without math null, add 2 Ω additional error in 2-wire ohms function. The 100M and 1G ohm ranges are 2-wire only.
[4]. This specification applies to the voltage measured from input terminal. 1 mA test current is the typical value. The change of current source leads to the variation in buck of diode junction.
[5]. AC Specification: It will be available after 60 minutes of warm-up, sine wave as well as 1/s speed rate.
[6]. Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50 kHz to 100 kHz, add 0.13% of range. The measurement range of 400 VAC is limited within the range of 7.5 x 10^7 Volt–Hz.
[7]. Three speed settings provided for low-frequency performance: 1/s (3 Hz), 5/s (20 Hz), 20/s (200 Hz). Additional errors will Not occur for the frequency greater than the filter settings.
[8]. Specifications are for sinewave input >5% of range, and is beyond 10 μA AC. For inputs from 1% to 5% of range, add 0.1% of range additional error.
[9]. This specification will be available after 60 minutes of warm-up and sine wave input, unless stated otherwise. This specification applies to 1s gate time.
[10]. This specification is available when both sine wave and square wave input ≥ 100 mV. For the input of 10 mV to 100 mV, the % of reading error needs to be multiplied by 10 times.
[11]. The amplitude range is from 10% to 120% and is lower than 400 VAC.
[12]. The input ≥ 60 mV, for 300 k ~ 1 MHz, within 100mV range.
[13]. The actual measurement range and test lead error will be constrained by the adopted test lead. The test lead accuracy adder covers all errors of measurements and ITS-90 temperature change.
[14]. Specifications are for film Capacitance inputs that are greater than 10% range. range.


Do you have a Data Acquisition System (Mainframe) with GPIB that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Data Acquisition System (Mainframe) with GPIB calibrated today! Click Here!

 

There are currently no reviews for this product.

 

Review the GW Instek DAQ-9600G - Data Acquisition System (Mainframe) with GPIB

Data Acquisition System (Mainframe) with GPIB

Name :

Email :

Title :