Time Electronics 7050 - Multifunction Process Calibrator
The 7050 Process Calibrator combines the essential functions of measurement and simulation of volts, millivolts, milliamps, and ohms with the direct readout simulation of thermocouples and RTD’s in °C or °F units.The instrument is microprocessor based and housed in a rugged case designed for use in the field. Input and Output functions are displayed simultaneously and the switch controls allow for easy and precise operation.
Five calibration points are instantaneously accessible through a preset percentage selector and all functions may be generated in a fully programmable ramp format.In addition to the standard calibration facilities the instrument may be used as a signal converter. Any electronic input signal can be converted and output as a proportional signal of 4-20mA, 0-5V or sq. root 4-20mA.
The 7050 is highly adaptable, IPT 68 and ITS 90 curves are accommodated and non-standard thermocouples and RTD’s may be programmed on request.
- Measure to 40V, 400mA, 40kΩ
- Output voltage up to 10V
- Output current to 20mA
- Output resistance to 40kΩ
- Accuracy 0.03% of range
- Thermocouple measure & simulate
- RTD measure & simulate: PT100
Technical Specifications
Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| 40V | 1mV | 0.03% |
| 4V | 100μV | 0.03% |
| 400mV | 10µV | 0.03% |
| 40mV | 1µV | 0.03% |
Current
| Range | Resolution | Accuracy |
|---|---|---|
| 400mA | 10μA | 0.03% |
| 40mA | 1μA | 0.03% |
| 4-20mA/0-100% | 0.01% | 0.06% |
Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 40000Ω | 1Ω | 0.03% |
| 4000Ω | 0.1Ω | 0.03% |
| 400Ω | 0.01Ω | 0.06% |
Output
Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| -2 to +10V | 1mV | 0.06% |
| -100mV to +400mV | 10μV | 0.03% |
| -10mV to +40mV | 1µV | 0.03% |
Current
| Range | Resolution | Accuracy |
|---|---|---|
| 0 to 20mA | 10μA | 0.03% |
| 4 to 20mA | 1μA | 0.03% |
| TX Sim (4 to 20mA) | 0.01% | 0.03% |
Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 0 to 400Ω | 0.01Ω | ± 0.03% |
Excitation Current 0.2 to 2mA
Other Specifications: Loop drive supply: Nominal 24V, current limit set to 30mA
Process Signal and Temperature Measurement & Simulation
Temperature
| Thermocouple Type | Temp Range °C | Accuracy °C |
|---|---|---|
|
J |
-210 to 0 |
± 0.4 |
|
K |
-270 to -240 |
± 3.0 |
|
T |
-270 to -250 |
± 3.0 |
|
R |
-50 to 0 |
± 3.0 |
|
S |
-50 to 0 |
± 2.5 |
|
B |
200 to 400 |
± 5.0 |
|
N |
0 to 330 |
± 0.4 |
|
E |
-230 to 0 |
± 0.5 |
RTD-PT100
| Range °C | Accuracy °C |
|---|---|
| -200 to 850 | ± 0.3 |
| Thermocouple | Linearisation to BS 4937. ITS 90 or IPTS68 selectable. |
| Pt100 | Linearisation to BS1904 (1984).DIN43760 (1980). Linearisation Acc: 0.1% of reading for Class A, 38.5Ω F.I |
| Excitation current | 0.5mA (Input), 200μA to 2mA (Output). |
| CJC | External (Pt100), Internal, Ice point. Accuracy +/-0.2°C at 23°C |
| Ramp | Fully programmable continuous and stepped |
| Steps | 5 Steps with fully adjustable zero |
| Transmitter | Tx function 4-20mA |
| Signal Converter | Any input to any output |
| Dual Readout | Measure and source simultaneously displayed |
General Specifications
| Stability | 100ppm per °C range 0 to 40°C |
| Displays | LCD Dot Matrix, 1 line by 20 charx12mm, alphanumeric. |
|
Dimensions / Weight |
W270 x H175 x D250mm, weight 4.5kg |
| Optional Extras | Calibration Certificates traceable to NPL and UKAS |
Do you have a Multifunction Process Calibrator that needs to be calibrated?
We offer standard traceable calibration and ISO 17025 accredited calibration services.
Get your Multifunction Process Calibrator calibrated today! Click Here!
There are currently no reviews for this product.