Your Cart
(empty)
0
Tradeport Electronics
measurement is our business
Cart : (empty)

Time Electronics 7015 - Dual Channel Pressure Calibrator

 Dual Channel Pressure Calibrator

Time Electronics

Rated 0 / 5 (0 reviews)

Category : Portable Process & Pressure Instruments

by quote only add to cart

Documents:
Datasheet

Overview
Features
Specifications
Calibration

The 7015 is a fully integrated pressure test and calibration system. It is suitable for field and laboratory calibration work of both pressure and electrical process loop signals. The instrument has two pressure channels with the ranges specified on ordering. Standard available ranges for each channel are; 0.2, 2, 5, 10, 20, 35, 70, 100, 200, 400, and 600bar. Vacuum down to -850mbar is possible on the 2bar version.

Best accuracy is 0.04% of range and maximum display resolution is 0.005% of range. The excellent resolution allows higher accuracies to be achieved under controlled lab conditions when accurate pressure standards such as dead-weight testers are available for transfer calibration.

For process signal calibration the loop signal is displayed on the LCD display to 0.05% accuracy. Loop drive supply (24V or 36V) is also provided. Both pressure and loop signal can be displayed at the same time to speed up the calibration of pressure transmitters. Additionally a continuity function is available for testing for open circuit loops.

The 7015 allows switch selection of five pressure units on channel one and four pressure units on channel two. These must be specified on ordering from the following: bar, PSI, kPa, MPa, inWg, cmWg, inHg, mmHg, Kg/cm2, atm. Both pressure and loop current are displayed on 4.5 digit LCD displays.

Options

For the 7015, input pressure can be generated by external hand pumps. The 7090 pneumatic (-950mbar to 40bar), or 7095 hydraulic (0 to 700bar) pumps are ideal for this purpose. Gauges are also available.

For process control signal simulation Time Electronics’ handheld 7006 Loop-Mate is available as an option. It provides 4 - 20mA and 0 - 10V signals at levels of 0%, 10%, 25%, 50%, 75%, 90%, 100%. It can also step through these set points automatically to allow hands free calibration. It is battery powered and can be stored in the 7015’s supplied carry pouch.

The 7015 is fitted with RS232 connection as standard. Display readback can be performed using EasyCal calibration software. With comprehensive features EasyCal can simplify and speed up calibration procedures. Easily produce calibration certificates and reports to international quality standards.

Portable Operation

Power is via internal rechargeable cells (24 hours typical use) with a built-in mains charger. The unit can be powered directly from the mains if required. The 7015 is housed in an impact resistant safety case that provides protection and increased durability for field use. All pressure connections are on the side of the case. The unit is supplied with 2 pressure hoses, electrical test lead, 2 electrical probes and 2 crocodile clips.

 

  • Vacuum to 600bar versions
  • Accuracy 0.04% of full scale
  • Pneumatic or hydraulic
  • Over- pressure alarm
  • Isolated 24/36V loop supply
  • Loop current measure
  • Mains / battery power

 

Technical Specifications

Pressure

Range (bar) Vac 0.2 2 5 10 20 35 70 100 200 400 600

Accuracy (% of fs)

0.04

0.1

0.04

0.04

0.04

0.04

0.04

0.04

0.04

0.1

0.1

0.1

Resolution (bar)

.0001

0.01mb

.0001

.001

.001

.001

0.01

0.01

0.01

0.01

0.1

0.1

Temperature Stability Less than 70ppm per °C

Units

bar, PSI, kPa, MPa, inWg, cmWg, inHg, mmHg, Kg/cm2, atm (5 units for channel 1, 4 for channel 2)

Maximum Pressure 2 x range, 1.5 x range for 20bar and above
Sensor Piezo-resistive - stainless steel diaphragm
Over pressure warning 1.2 x range full scale - audio and visual (on LCD) warning
Fittings Minimess (>20bar versions) & Quick Release (<20bar versions)

Electrical

Range Loop current measurement 0 to 200mA
Resolution 10μA
Resistance Loop load 5Ω
Accuracy 0.05% of reading ±1 digit
Loop Drive 24V or 36V switch selectable, 50mA max - isolated and with short circuit protection
Continuity Threshold: 100Ω with audio and visual warning
Terminals 4mm industry standard terminal posts

General Specifications

Power Source Internal rechargeable NiMH batteries, or mains supply (230V or 110V 50/60Hz)
Battery Life 24 hrs of typical usage between charges
Display 7 segment LCD with ‘Low Battery’ warning indicator
Interfaces RS-232 for readback and re-calibration (software not supplied)
Case Structural resin which is weather-proof to IP66 standard
Dimensions W270 x H175 x D250mm
Weight 3kg

Supplied with

2 pressure hoses: Minimess for versions ordered over 20bar, Quick release for under 20bar. Electrical Test Lead: Twin flexible lead, terminated with 4 off 4mm gold plated plugs. Electrical probes: 2 off low voltage probes and 2 crocodile clips. Mains power lead. 

Optional Extras Pneumatic & hydraulic hand pumps, gauges, loop simulators, EasyCal calibration software, spare fuses, and calibration certificates traceable to NPL and UKAS
Country of Origin UK

Do you have a Dual Channel Pressure Calibrator that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.