Time Electronics 5018 - Programmable DC/AC V/I Calibrator

Programmable DC/AC V/I Calibrator

Time Electronics

Rated 0 / 5 (0 reviews)

Category : Multifunction Calibrators

by quote only add to cart

Documents:
Datasheet

Overview
Features
Specifications
Calibration

High Accuracy Programmable DC/AC Source

The 5018 is a versatile instrument that covers a wide range of applications. It can be configured as a simple benchtop DC voltage source or advanced AC/DC voltage and current calibrator controlled via PC or laptop, performing any number of tasks as part of a complex ATE test rig.

Ramping Feature

The 5018 includes an internal ramping feature. A ramp rate per range for each function can be set via a PC. The option to ramp the output can be turned on or off via the front panel. This is particularly useful in aviation and automotive industries for testing analogue gauges.

Simple Operation and Clear Visual Display

Simple front panel operation allows the user to quickly set the function and output required. Using the jog / shuttle dial deviation the user can finely adjust the output value as a percentage (+/- 99.99%). All this information is shown on a clear, easy to read LED display.

GPIB, RS-232, and USB Interfaces

These interfaces allow the 5018 to be connected to a PC and controlled by an external program such as Time Electronics’ EasyCal calibration software. The 5018 uses a SCPI command structure for programming. This makes writing control programs in Visual Basic, C and Labview a simple task.

EasyCal Software - Manage, Automate and Optimise the Calibration Process

Connect the 5018 to a PC or laptop (via RS-232, GPIB or USB) installed with Time Electronics EasyCal software and automate the calibration process. Increase speed of calibration and consistency of results. Easily produce calibration certificates and reports to ISO 9001, ISO 17025, and other international quality standards.

EasyCal has the ability to control a wide range of programmable instruments and comprehensive ATE systems can be configured to control a variety of applications. It has full read back capability and therefore can perform closed loop testing. Conditional tests can be configured to allow decision making to further enhance the usefulness.

 

  • 0 - 22V DCV (30V DC optional)
  • 0 - 22V AC, 220V & 1kV DC options
  • Optional 0 - 220mA AC/DC
  • Ramping feature
  • GPIB / RS-232 / USB interfaces
  • Ideal for ATE applications
  • EasyCal Software compatible

 

Technical Specifications

DC Voltage (30V Options 9724 Shown in Brackets)

Range Accuracy 1 year Output Resistance Max Output Current Resolution
22mV (33mV) 30ppm + 4μV 10Ω (50Ω) - 100nV
220mV (330mV) 15ppm (30ppm) + 6μV 10Ω (50Ω) - 1μV
2.2V (3.3V) 15ppm (20ppm)+ 20μV < 0.15Ω 20mA 1μV
22V (30V) 15ppm (20ppm) + 150μV < 0.15Ω 20mA 10μV

Options

DC High Voltage Options 9720 / 9721 (9721 Includes 9720 Range)

Range Accuracy 1 year Output Resistance Max Output Current Resolution
220V* 15ppm + 1mV <0.25Ω 10mA 100μV
1kV* 25ppm + 3mV <1Ω 1mA 1mV

*Minimum Load 20kΩ

DC Current Option 9718

Range Accuracy 1 year Compliance Voltage Resolution
220μA 150ppm + 15nA 11V 1nA
2.2mA 100ppm + 40nA 11V 10nA
22mA 80ppm + 200nA 11V 10nA
220mA 80ppm + 3μA 11V 100nA

AC V Option 9719 10Hz to 20KHz (Sine 0.01% Crystal Controlled)

Range RMS Acc % 1yr 10Hz - 1kHz Acc % 1 yr 1kHz - 10kHz Acc % 1yr 10kHz - 20kHz Output Resistance Max Output Current
22mV 0.05% + 100μV 0.05% + 150μV 0.05% + 250μV 10Ω -
220mV 0.04% + 100μV 0.04% + 150μV 0.04% + 250μV 10Ω -
2.2V 0.03% + 170μV 0.03% + 250μV 0.03% + 300μV <0.15Ω 20mA
22V 0.03% + 2.0mV 0.03% + 3.0mV 0.03% + 4.0mV <0.15Ω 20mA

AC Current Option 20Hz to 1KHz (Sine) - Available if Current and AC Options Are Fitted

Range RMS Accuracy 1 year Compliance Voltage RMS Resolution
220μA 0.07% + 300nA 5V 10nA
2.2mA 0.05% + 300nA 5V 10nA
22mA 0.05% + 3μA 5V 100nA
220mA 0.05% + 30μA 5V 1μA

General Specifications

Warm up 1 hour to full accuracy
Settling Time Less than 5 seconds
Standard Interfaces GPIB (IEEE-488), RS-232, USB
Temp Performance Operating: 10 to 40°C, Full Spec: 23°C +/- 5°C, Storage: -10°C to 50°C
Operating Humidity < 80% non condensing
Operating Altitude 0 to 3km. Non operating 3 to 12km
Line Power 100 - 230V AC 50/60 Hz. Power Consumption 60W typical, 80W Max.
Dimensions W450 x D272 x H152mm (18 x 11 x 7”)
Weight 8.2Kg (18lbs)
Country of Origin UK

Supplied With

User manual, RS-232 cable, and USB adaptor/cable


Do you have a Programmable DC/AC V/I Calibrator that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Programmable DC/AC V/I Calibrator calibrated today! Click Here!

 

There are currently no reviews for this product.

 

Review the Time Electronics 5018 - Programmable DC/AC V/I Calibrator

Programmable DC/AC V/I Calibrator

Name :

Email :

Title :

Review: