Fluke 714B - Thermocouple Calibrator
Overview
Features
Specifications
Calibration
For the temperature calibration professional that wants a highly accurate, easy-to-use, single function thermocouple temperature calibrator the Fluke 714B is the ideal test tool.
- The 714B can measure and simulate (17) different thermocouple types and millivolts
- Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Hanging tool designed in and included with every unit
- Configurable 0% and 100% source settings for quick 25% linearity checks
- Linear ramp and 25% step auto ramp based on 0% and 100% settings
- Dual inputs and backlit display for easy interpretation of measurements
- Power down settings remembered at power up for easy restart of tests
- 1-year and 2-year specifications and traceable certification of calibration
| General Specifications | |||||
| Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V | ||||
| Operating temperature | -10°C to 50°C | ||||
| Storage temperature | -30°C to 60°C | ||||
| Operating altitude | 2,000 m | ||||
| Storage altitude | 12,000 m | ||||
| Relative humidity (% RH operating without condensation) | Non condensing 90% (10°C to 30°C) 75% (30°C to 40°C) 45% (40°C to 50°C) (Without condensation) |
||||
| Vibration requirements | MIL-T-28800E, Class 2 | ||||
| Drop text requirements | 1 m | ||||
| IP Rating | IEC 60529: IP52 | ||||
| Electromagnetic environment | IEC 61326-1, Portable | ||||
| Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 | ||||
| Power supply | 4 AA NEDA 1.5A IEC LR6 batteries | ||||
| Size (H x W x L) | 52.5 x 84 x 188.5 mm | ||||
| Weight | 515 g | ||||
| DC mA Measurement | |||||
| Resolution | Range | Accuracy (% of reading + counts) | |||
| 0-24 mA | 0.001 mA | 0.010% + 2 μA | |||
| Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) | ||||
| Millivolt Measurement and Source | |||||
| Resolution | Range | Accuracy (% of reading +counts) | |||
| -10 mV to 75 mV | 0.01 mV | 0.015% + 10 μA | |||
| Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) | ||||
| Thermocouple Input and Output | |||||
| TC Type | Range (°C) | Measure (°C) | Source (°C) | ||
| 1 year | 2 year | 1 year | 2 year | ||
| E | -250 to 200°C | 1.3 | 2.0 | 0.6 | 0.9 |
| -200 to -100°C | 0.5 | 0.8 | 0.3 | 0.4 | |
| -100 to 600°C | 0.3 | 0.4 | 0.3 | 0.4 | |
| 600 to 1000°C | 0.4 | 0.6 | 0.2 | 0.3 | |
| N | -200 to -100°C | 1.0 | 1.5 | 0.6 | 0.9 |
| -100 to 900°C | 0.5 | 0.8 | 0.5 | 0.8 | |
| 900 to 1300°C | 0.6 | 0.9 | 0.3 | 0.4 | |
| J | -210 to -100°C | 0.6 | 0.9 | 0.3 | 0.4 |
| -100 to 800°C | 0.3 | 0.4 | 0.2 | 0.3 | |
| 800 to 1200°C | 0.5 | 0.8 | 0.3 | 0.3 | |
| K | -200 to -100°C | 0.7 | 1.0 | 0.4 | 0.6 |
| -100 to 400°C | 0.3 | 0.4 | 0.3 | 0.4 | |
| 400 to 1200°C | 0.5 | 0.8 | 0.3 | 0.4 | |
| 1200 to 1372°C | 0.7 | 1.0 | 0.3 | 0.4 | |
| T | -250 to -200°C | 1.7 | 2.5 | 0.9 | 1.4 |
| -200 to 0°C | 0.6 | 0.9 | 0.4 | 0.6 | |
| 0 to 400°C | 0.3 | 0.4 | 0.3 | 0.4 | |
| B | 600 to 800°C | 1.3 | 2.0 | 1.0 | 1.5 |
| 800 to 1000°C | 1.0 | 1.5 | 0.8 | 1.2 | |
| 1000 to 1820°C | 0.9 | 1.3 | 0.8 | 1.2 | |
| R | -20 to 0°C | 2.3 | 2.8 | 1.2 | 1.8 |
| 0 to 100°C | 1.5 | 2.2 | 1.1 | 1.7 | |
| 100 to 1767°C | 1.0 | 1.5 | 0.9 | 1.4 | |
| S | -20 to 0°C | 2.3 | 2.8 | 1.2 | 1.8 |
| 0 to 200°C | 1.5 | 2.1 | 1.1 | 1.7 | |
| 200 to 1400°C | 0.9 | 1.4 | 0.9 | 1.4 | |
| 1400 to 1767°C | 1.1 | 1.7 | 1.0 | 1.5 | |
| C | 0 to 800°C | 0.6 | 0.9 | 0.6 | 0.9 |
| 800 to 1200°C | 0.8 | 1.2 | 0.7 | 1.0 | |
| 1200 to 1800°C | 1.1 | 1.6 | 0.9 | 1.4 | |
| 1800 to 2316°C | 2.0 | 3.0 | 1.3 | 2.0 | |
| L | -200 to -100°C | 0.6 | 0.9 | 0.3 | 0.4 |
| -100 to 800°C | 0.3 | 0.4 | 0.2 | 0.3 | |
| 800 to 900°C | 0.5 | 0.8 | 0.2 | 0.3 | |
| U | -200 to 0°C | 0.6 | 0.9 | 0.4 | 0.6 |
| 0 to 600°C | 0.3 | 0.4 | 0.3 | 0.4 | |
| BP | 0 to 1000°C | 1.0 | 1.5 | 0.4 | 0.6 |
| 1000 to 2000°C | 1.6 | 2.4 | 0.6 | 0.9 | |
| 2000 to 2500°C | 2.0 | 3.0 | 0.8 | 1.2 | |
| XK | -200 to 300°C | 0.2 | 0.3 | 0.2 | 0.5 |
| 300 to 800°C | 0.4 | 0.6 | 0.3 | 0.6 | |
| G | 100 to 300°C | 1.6 | 2.4 | 1.2 | 1.8 |
| 300 to 1500°C | 1.0 | 1.5 | 1.0 | 1.5 | |
| 1500 to 2320°C | 2.0 | 3.0 | 1.6 | 2.4 | |
| D | 0 to 300°C | 1.6 | 2.4 | 1.2 | 1.8 |
| 300 to 1500°C | 1.0 | 1.5 | 1.0 | 1.5 | |
| 1500 to 2315°C | 2.0 | 3.0 | 1.6 | 2.4 | |
| P | 0 to 1000°C | 1.6 | 2.4 | 0.6 | 0.9 |
| 1000 to 1395°C | 2.0 | 3.0 | 0.8 | 1.2 | |
| M | -50 to 100°C | 1.0 | 1.5 | 0.4 | 0.6 |
| 100 to 1000°C | 1.6 | 2.4 | 0.6 | 0.9 | |
| 1000 to 1410°C | 2.0 | 3.0 | 0.8 | 1.2 | |
Do you have a Thermocouple Calibrator that needs to be calibrated?
We offer standard traceable calibration and ISO 17025 accredited calibration services.
Get your Thermocouple Calibrator calibrated today! Click Here!