Fluke 750P03 - Pressure Module 0 to 1 (0 to 350 MBAR) (0 to 35 KPA)
A family of 48 pressure modules covers pressure calibrations from 0 to 1 in H20 to 10,000 psi (2.5 mBar to 690 bar).
Gage pressure modules have one pressure fitting and measure the pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. Each module is clearly labeled for range, overpressure and media compatibility. All modules include NPT, metric (BSP) and M20 adapters.
Precision Pressure Measurement for 75X and 720 Series Calibrators
The 750P Series Pressure Modules are the ideal pressure modules to enable gage, differential and absolute pressure measurement with Fluke 750 and 740 series DPCs and 725, 726 MPCs to measure pressure.
- 0.025 % reference uncertainty
- 6-month and 1-year specifications
- Temperature compensated 0 °C to 50 °C
- Digital communication to calibrators, no analog losses or errors
- Broad selection of ranges
- Gage, differential, dual range, absolute and vacuum measurement models
| Specifications | |||||
| Differential | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750P00 | 0 to 1 in H2O (0 to 2.5 mBar) | 30X | Dry Air | Dry Air | |
| 750P01 | 0 to 10 in H2O (0 to 25 mBar) | 3X | Dry Air | Dry Air | |
| 750P02 | 0 to 1 psi (0 to 70 mBar) | 3X | Dry Air | Dry Air | |
| 750P22 | 0 to 1 psi (0 to 70 mBar) | 3X | 316 SS | Dry Air | |
| 750P03 | 0 to 5 psi (0 to 350 mBar) | 3X | Dry Air | Dry Air | |
| 750P23 | 0 to 5 psi (0 to 350 mBar) | 4X | 316 SS | Dry Air | |
| 750P04 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | |
| 750P24 | 0 to 15 psi (0 to 1 bar) | 4X | 316 SS | Dry Air | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750P00 | ± 0.15 % | ± 0.3 % | ± 0.35 % | ± 0.25 % | ± 0.30 % |
| 750P01 | ± 0.1 % | ± 0.2 % | ± 0.3 % | ± 0.15 % | ± 0.25 % |
| 750P02 | ± 0.050 % | ± 0.1 % | ± 0.15 % | ± 0.075 % | ± 0.125 % |
| 750P22 | ± 0.050 % | ± 0.1 % | ± 0.15 % | ± 0.075 % | ± 0.125 % |
| 750P03 | ± 0.02 % | ± 0.04 % | ± 0.05 % | ± 0.035 % | ± 0.04 % |
| 750P23 | ± 0.02 % | ± 0.04 % | ± 0.05 % | ± 0.035 % | ± 0.04 % |
| 750P04 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P24 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| Gage | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750P05 | 0 to 30 psi (0 to 2 bar) | 4X | 316 SS | N/A | |
| 750P06 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | N/A | |
| 750P27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | N/A | |
| 750P07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | N/A | |
| 750P08 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | N/A | |
| 750P09 | 0 to 1500 psi (0 to 100 bar) | 3X | 316 SS | N/A | |
| 750P2000 | 0 to 2000 psi (0 to 140 bar) | 3X | 316 SS | N/A | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750P05 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P06 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P27 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P07 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P08 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P09 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P2000 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| High | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750P29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | N/A | |
| 750P30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | N/A | |
| 750P31 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | N/A | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750P29 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P30 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| 750P31 | ± 0.0175 % | ± 0.035 % | ± 0.045 % | ± 0.03 % | ± 0.04 % |
| Absolute | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750PA3 | 0 to 5 psia (0 to 350 mBar) | 4X | 316 SS | N/A | |
| 750PA4 | 0 to 15 psia (0 to 1 bar) | 4X | 316 SS | N/A | |
| 750PA5 | 0 to 30 psia (0 to 2 bar) | 4X | 316 SS | N/A | |
| 750PA6 | 0 to 100 psia (0 to 7 bar) | 4X | 316 SS | N/A | |
| 750PA27 | 0 to 300 psia (0 to 20 bar) | 4X | 316 SS | N/A | |
| 750PA7 | 0 to 500 psia (0 to 35 bar) | 4X | 316 SS | N/A | |
| 750PA8 | 0 to 1000 psia (0 to 70 bar) | 3X | 316 SS | N/A | |
| 750PA9 | 0 to 1500 psia (0 to 100 bar) | 3X | 316 SS | N/A | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750PA3 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA4 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA5 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA6 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA27 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA7 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA8 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| 750PA9 | ± 0.03 % | ± 0.06 % | ± 0.07 % | ± 0.05 % | ± 0.06 % |
| Vacuum | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750PV3 | -5 psi (-350 mBar) | 4X | 316 SS | Dry Air | |
| 750PV4 | -15 psi (-1 bar) | 4X | 316 SS | Dry Air | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750PV3 | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
| 750PV4 | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
| Dual | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750PD2 | -1 to 1 psi (-70 to 70 mBar) | 4X | 316 SS | Dry Air | |
| 750PD3 | -5 to 5 psi (-350 to 350 mBar) | 4X | 316 SS | Dry Air | |
| 750PD10 | -10 to 10 psi (-700 to 700 mBar) | 4X | 316 SS | Dry Air | |
| 750PD4 | -15 to 15 psi (-1 to 1 bar) | 4X | 316 SS | Dry Air | |
| 750PD5 | -15 to 30 psi (-1 to 2 bar) | 4X | 316 SS | — | |
| 750PD50 | -15 to 50 psi (-1 to 3.5 bar) | 4X | 316 SS | — | |
| 750PD6 | -15 to 100 psi (-1 to 7 bar) | 4X | 316 SS | — | |
| 750PD7 | -15 to 200 psi (-1 to 14 bar) | 4X | 316 SS | — | |
| 750PD27 | -15 to 300 psi (-1 to 20 bar) | 4X | 316 SS | — | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750PD2 | ± 0.05% | ± 0.1% | ± 0.15% | ± 0.075% | ± 0.125% |
| 750PD3 | ± 0.03% | ± 0.06% | ± 0.07% | ± 0.05% | ± 0.06% |
| 750PD10 | ± 0.025% | ± 0.05% | ± 0.07% | ± 0.04% | ± 0.06% |
| 750PD4 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| 750PD5 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| 750PD50 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| 750PD6 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| 750PD7 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| 750PD27 | ± 0.0175% | ± 0.035% | ± 0.045% | ± 0.03% | ± 0.04% |
| Reference | |||||
| Model | Parameter / Range | Burst Rating6 | Hi Side Media2 | Lo Side Media2 | |
| 750R045 | 0 to 15 psi (0 to 1 bar) | 3X | Dry Air | Dry Air | |
| 750R065 | 0 to 100 psi (0 to 7 bar) | 4X | 316 SS | — | |
| 750R27 | 0 to 300 psi (0 to 20 bar) | 4X | 316 SS | — | |
| 750R07 | 0 to 500 psi (0 to 35 bar) | 4X | 316 SS | — | |
| 750R085 | 0 to 1000 psi (0 to 70 bar) | 3X | 316 SS | — | |
| 750R29 | 0 to 3000 psi (0 to 200 bar) | 3X | 316 SS | — | |
| 750R30 | 0 to 5000 psi (0 to 340 bar) | 3X | 316 SS | — | |
| 750R315 | 0 to 10000 psi (0 to 700 bar) | 2X | 316 SS | — | |
| 750RD5 | -15 to 30 psi (-1 to 2 bar) | 4X | Dry Air | — | |
| 750RD65 | -12 to 100 psi (-1 to 7 bar) | 4X | 316 SS | — | |
| 750RD27 | -12 to 300 psi (-0.8 to 20 bar) | 4X | 316 SS | — | |
| Model | Reference Uncertainty4 | Total Uncertainty 1-year (15-35 °C) |
Total Uncertainty 1-year1 |
Total Uncertainty 6-month (15-35 °C) |
Total Uncertainty 6-month1 |
| 750R045 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R065 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R27 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R07 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R085 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R29 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R30 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750R315 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750RD5 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750RD65 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 750RD27 | ± 0.01% of FS | ± 0.02% of FS | ± 0.04% of FS | ± 0.015% of FS | ± 0.035% of FS |
| 1. Total uncertainty, % of full span for temperature range 0 °C to +50 °C, one year interval. Total uncertainty, 1.0% of full span for temperature range -10 °C to 0 °C, one year interval. No 6 month specification available for range -10 °C to 0 °C. 2. "Non-corrosive gasses" indicates dry air or non-corrosive gas as compatible media. "Stainless Steel 316-SS" indicates media compatible with Type 316 Stainless Steel. 3. Specifications % of Full Span unless otherwise noted. 4. Reference Uncertainty is the specification for as left data for 24 hours. 5. When reference class modules are used with fixed resolution products (717, 718, 719 series, 725 and 726) calibrators add ± 1 count to the overall accuracy specification. 6. Burst rating specification refers to the multiplier times full scale of the module for the rated burst pressure. |
|||||
Do you have a Pressure Module 0 to 1 (0 to 350 MBAR) (0 to 35 KPA) that needs to be calibrated?
We offer standard traceable calibration and ISO 17025 accredited calibration services.
Get your Pressure Module 0 to 1 (0 to 350 MBAR) (0 to 35 KPA) calibrated today! Click Here!
There are currently no reviews for this product.