Soken DAC-ASC-5 - C & tanδ Meter
DAC-ASC-5 realizes measurement in 1 second or less to save your testing time with the best accuracy and repeatability. DAC-ASC-5 is compact in size and weight, and provided with GP-IB interface as standard to control with PC. Thus, it is applicable for automating testing system in wide ranges. Moreover, the conventional judging function for capacitance error and tanδ can be set through GP-IB interface, and is allowed to set a separate judging range in accordance with the testing condition of each test object. DAC-ASC-5 can assist in productivity of your inspection line.
- Measuring Range : 0.01μF - 200μF or 1100μF or 3300μF
- Measuring time : 1.0 second or less
- GP-IB is provided, all programmable and controlled by PC
200uF / 600V type
| Test Voltage | AC50V - 600V |
| Test Range | Capacitance 0.01uF - 200uF tanδ 0~20% |
| Resolution | ΔC 0.01% tanδ'0.001% |
| Accuracy | ΔC ±(0.2%+3%Rdg+2digit) tanδ ±(0.01%+3%Rdg+2digit) |
| Measuring Time | 1 second or less *2 |
| Interface | GP-IB (IEEE-488-1987 Compliance) |
| AC Source | AC100V-240V±10%, 50/60Hz |
| Component Units | Main Unit: W430×H199×D380mm, 20kg |
| Composition type | A |
1100uF / 600V type *1
| Test Voltage | AC50V - 600V |
| Test Range | Capacitance 0.01uF - 1100uF tanδ 0~20% |
| Resolution | ΔC 0.01% tanδ'0.001% |
| Accuracy | ΔC ±(0.2%+3%Rdg+2digit) tanδ ±(0.01%+3%Rdg+2digit) |
| Measuring Time | 1 second or less *2 |
| Interface | GP-IB (IEEE-488-1987 Compliance) |
| AC Source | AC100V-240V±10%, 50/60Hz |
| Component Units | Main Unit: W430xH199xD380mm, 18kg CT&Cs Unit: W210xH240xD200mm, 5kg Switch Unit: W210xH250xD200mm, 8kg |
| Composition type | B |
*1: Maximum testing current: 200A
*2: W hen measuring such specimen as Δ C>20% or tanδ >10%, the measuring time become longer.
1100uF / 1200V type *1
| Test Voltage | AC100V - 1200V |
| Test Range | Capacitance 0.01uF - 1100uF tanδ 0~20% |
| Resolution | ΔC 0.01% tanδ ±0.001% |
| Accuracy | ΔC ±(0.2%+3Rdg+2digit) tanδ ±(0.01%+3Rdg+2digit) |
| Measuring Time | 1 second or less *2 |
| Interface | GP-IB (IEEE-488-1987 Compliance) |
| AC Source | AC100V-240V ±10%, 50/60Hz |
| Component Units | Main Unit: W430 x H199 x D380mm, 18kg CT&Cs Unit: W210 x H240 x D200mm, 5kg Switch Unit: W210 x H250 x D200mm, 8kg |
3300uF / 1200V type *1
| Test Voltage | AC100V - 1200V |
| Test Range | Capacitance 0.01uF - 3300uF tanδ 0~20% |
| Resolution | ΔC 0.01% tanδ ±0.001% |
| Accuracy | ΔC ±(0.2%+3Rdg+2digit) tanδ ±(0.01%+3Rdg+2digit) |
| Measuring Time | 1 second or less *2 |
| Interface | GP-IB (IEEE-488-1987 Compliance) |
| AC Source | AC100V-240V±10%, 50/60Hz |
| Component Units | Main Unit: W430 x H199 x D380mm, 18kg CT&Cs Unit: W210 x H240 x D200mm, 5kg Switch Unit: W210 x H250 x D200mm, 8kg |
*1: Maximum testing current: 200A
*2: W hen measuring such specimen as Δ C>20% or tanδ >10%, the measuring time become longer.