Soken DAC-PBV-8 - Electronic Voltage Burden
Overview
Features
Specifications
Calibration
- Programmable burden used for voltage error testing in accordance with IEC60044-2.
- Testing voltage can be varied over a wide range: 2 % to 190% of the rated secondary voltage.
- Burdens from 0 to 50VA are available standalone, and a maximum of 400VA in steps of 100VA is available with optional electronic burdens.
- USB Interface is a standard fixture.
- Test conditions can be easily configured: the maximum of 100 test parameters (rating, burden, power factor, and lead wire compensation) are programmable.
Option Electric Burden
Optional burden DAC-FBV series are available in need of further VA. The burdens are controlled by DAC-PBV-8 through serial communication.
DAC-FBV Series: 50VA, 100VA, and 200VA type. When combining all units, max 400VA is available.
Connecting Diagram

- Programmable burden used for voltage error testing in accordance with IEC60044-2.
- Testing voltage can be varied over a wide range: 2 % to 190% of the rated secondary voltage.
- Burdens from 0 to 50VA are available standalone, and a maximum of 400VA in steps of 100VA is available with optional electronic burdens.
- USB Interface is a standard fixture.
- Test conditions can be easily configured: the maximum of 100 test parameters (rating, burden, power factor, and lead wire compensation) are programmable.
General
| Rated Secondary voltage | 100, 110, 115, 120, 200, 230V and each x 1/√ 3, and 100/3, 110/3, 200/3V | ||||||
| Test voltage range | 2% to 190% of the rated secondary voltage (Max current of 12A) | ||||||
| Rated burden setting | 0.00 to 50.0 VA Maximum of 400VA* with optional Electric burden *Except rated secondary voltages:200, 230, 200/√ 3 and 230/√ 3 | ||||||
| Burden Power factor |
| ||||||
| Test Frequency | 50/60 Hz | ||||||
| Settable resolution |
| ||||||
| Burden Power Factor | 0.01 steps | ||||||
| Accuracy (target value) | Burden and Burden Power Factor: ±4% | ||||||
| Communication | USB (2.0/1.1) | ||||||
| Other Functions | Test condition programming, over current protection, Alarm function for measuring circuit | ||||||
| Rated Supply Voltage | 100 to 240V ±10% | ||||||
| Dimensions | W430 x H250 x D380mm (Excluding potions) |
Do you have a Electronic Voltage Burden that needs to be calibrated?
We offer standard traceable calibration and ISO 17025 accredited calibration services.
Get your Electronic Voltage Burden calibrated today! Click Here!