GW Instek GPM-8340 - Introduction of GPM-8340 4-Element AC Power Meter

Introduction of GPM-8340 4-Element AC Power Meter

GW Instek

Rated 0 / 5 (0 reviews)

Category : AC Power Meter

by quote only add to cart

Documents:

Overview
Features
Specifications
Calibration

▪ Introduction of GPM-8340 4-Element AC Power Meter

The GW Instek GPM-8340 is a digital power meter designed specifically for multiphase AC power measurements, making it suitable for a wide range of electrical and electronic product testing applications. Featuring a test bandwidth of DC, 0.1 Hz to 10 kHz, a 16-bit A/D converter, and a 300 kHz sampling rate, the GPM-8340 utilizes a 5” TFT LCD display, 5-digit measurement display, provides 28 relevant power measurement parameters, and offers high-precision measurement capabilities. It also features waveform display (voltage/current/power), integration measurement function, harmonic measurement and analysis (50 Hz / 60Hz measurement compliant with IEC61000-4-7 requirements), external sensor input terminals, and various communication interfaces, helping users achieve clear, convenient, and accurate power measurements. It is considered one of the most feature-rich and cost-effective power meters in its class.

Product Features

  • ※ 4 Input Elements
  • ※ 5” TFT LCD
  • ※ Voltage/Current Test Freq. Bandwidth: DC, 0.1 Hz to 10 kHz
  • ※ Lowest current range 5 mA & 0.1 µW watt resolution
  • ※ Waveform Display: V (Voltage), I (Current), P (Wattage)
  • ※ Meeting IEC 61000-4-7 harmonic measurement (50 Hz/ 60 Hz)
  • ※ Wiring selecting button (1P2W, 1P3W, 3P3W, 3P4W, 3V3A)
  • ※ Harmonic measurement & analysis up to 50 orders
  • ※ Supports inrush current and energy measurement
  • ※ Provide external current sensor input (EXT)
  • ※ Standard Interface: RS-232C, USB device/ host, LAN, Digital I/O and mini GPIB or CAN (select one)
GENERAL
⚠ Note The below are the basic conditions required to operate the GPM-8340 within specifications:
Calibration: Yearly
Operating environment: (18 to 28) °C (64.4 to 82.4) °F
Humidity: < 80 % RH,
Accuracy: ± (% of reading + % of range)
The specifications apply when it warmed up for at least 30 min and operates in the slow rate.
The power supply cable must be grounded to ensure accuracy.
Input voltage and current must be standard sine wave.
The power factor must be 1.
The common-mode voltage must be zero.
Specification Condition Temperature: 23 °C ± 5 °C; Humidity: < 80 %RH (non-condensing)
Operation Condition Temperature 0 °C to 40 °C
30 °C to 40 °C, Relative Humidity < 70 %RH (non-condensing)
> 40 °C, Relative Humidity < 50 %RH (non-condensing)
Indoor use only
Altitude: < 2000 m
Pollution degree 2
Storage Condition Temperature −10 °C to 70 °C; Humidity: < 90 %RH (non-condensing)
Power Source AC 100 V to 240 V, 50 Hz to 60 Hz ; Consumption Max. 140 VA
Dimensions 220 mm(W) x 132 mm(H) x 444 mm(D)
Weight Approx. 5.24 kg
INPUT
Item Specifications
Input Type Voltage Floating input through resistive voltage divider
Current Floating input through shunt
Measure Range Voltage (CF=3) 15 V, 30 V, 60 V, 150 V, 300 V, 600 V and 1000 V
Current (CF=4)
Direct input 5 mA, 20 mA, 50 mA, 200 mA, 500 mA, 2 A, 5 A, 20 A
Sensor input EXT: 10 mV, 25 mV, 50 mV, 100 mV, 2.5 V, 5 V, 10 V
Input Impedance Voltage Input resistance: approach 4 MΩ
Current
Direct input range (5 to 200) mA Input resistance: approach 505 mΩ
Direct input range (0.5 to 20) A Input resistance: approach 5 mΩ
Sensor input Input resistance: approach 100 kΩ
Continuous Maximum
Allowable Input
Voltage Peak value of 1.5 kV or RMS value of 1 kV, whichever is less.
Current
Direct input range (5 to 200) mA Peak value of 30 A or RMS value of 20 A, whichever is less.
Direct input range (0.5 to 20) A Peak value of 100 A or RMS value of 30 A, whichever is
Sensor input Peak value less than or equal to 5 times of the rated range
Input Bandwidth DC, 0.1 Hz to 10 kHz
Continuous Maximum
Common-mode Voltage
600 Vrms, CAT Ⅱ
Line Filter select OFF or ON (cut off frequency of 5.5 kHz)
Frequency Filter select OFF or ON (cut off frequency of 5.5 kHz)
A/D Converter Simultaneous conversion voltage and current inputs
Resolution 16 bits
Maximum conversion rate Approx. 300 kHz
VOLTAGE AND CURRENT ACCURACY
Item Specifications
Requirements Temperature 23 °C ± 5 °C
Humidity 30 % to 75 %RH
Input waveform Sine wave
common-mode voltage 0 V
Number of displayed digits 5 digits
Frequency filter Turn on to measure voltage or current of 200 Hz or less
After 30 minutes after warm-up time has passed
After measurement range is changed (zero-level compensation)
Update interval is 250 ms
Accuracy Effective range 1 % to 105 % of range
DC ± (0.1 % of reading + 0.2 % of range)
0.1 Hz ≤ f < 45 Hz ± (0.1 % of reading + 0.2 % of range)
45 Hz ≤ f ≤ 66 Hz ± (0.1 % of reading + 0.05 % of range)
66 Hz < f ≤ 1 kHz ± (0.1 % of reading + 0.2 % of range)
1 kHz < f ≤ 10 kHz ± ((0.07 x f) % of reading + 0.3% of range)
Temperature Coefficient Add ± 0.03 % of reading/°C within the range 5 °C to 18 °C or 28 °C to 40 °C.
When the Line Filter is
Turned ON
45 Hz to 66 Hz Add 0.2 % of reading
< 45 Hz Add 0.5 % of reading
Accuracy Changes Caused by Data Update Interval When the data update interval is 100 ms, and Auto, add 0.05 % of reading to the 0.1 Hz to 1 kHz accuracy.
Influence of Temperature Changes After Zero-level Compensation or Range Change Add 0.02 % of range/°C to the DC voltage accuracy.
Add the following value to the DC current accuracies.
5 mA / 20 mA / 50 mA / 200 mA ranges 5 μA/°C
0.5 A / 2 A / 5 A / 20 A ranges 500 μA/°C
EXT: 2.5 V/ 5 V/ 10 V 1 mV/°C
EXT: 10 mV/ 25 mV/ 50 mV/ 100 mV 50 μV/°C
ACTIVE POWER ACCURACY
Item Specifications
Requirements Same as the conditions for voltage and current.
Power factor 1
Accuracy Effective range 1 % to 110 % of range
DC ± (0.1 % of reading + 0.2 % of range)
0.1 Hz ≤ f < 45 Hz ± (0.3 % of reading + 0.2 % of range)
45 Hz ≤ f ≤ 66 Hz ± (0.1 % of reading + 0.05 % of range)
66 Hz < f ≤ 1 kHz ± (0.2 % of reading + 0.2 % of range)
1 kHz < f ≤ 10 kHz ± (0.1 % of reading + 0.3 % of range) ± [{0.067 x (f-1)} % of reading]
Influence of Power Factor when power factor (λ) = 0 (S: apparent power)
± 0.1 % of S for 45 Hz ≤ f ≤ 66 Hz
± {(0.1 + 0.15 × f) % of S }
•f is frequency of input signal in kHz
when 0 < λ < 1 (Φ: phase angle of the Voltage and current)
(power reading ) × [(power reading error%) + (power range %) × (power range / indicated
When the Line Filter is
Turned ON
45 Hz to 66 Hz Add 0.3 % of reading
< 45 Hz Add 1 % of reading
Temperature Coefficient Same as the temperature coefficient for voltage and current
Accuracy of Apparent Power S Voltage accuracy + current accuracy
Accuracy of Reactive Power Q Accuracy of apparent power + (√1.0004 - λ2) - (√1 - λ2) × 100 %
Accuracy of Power Factor λ ± [(λ-λ/1.0002)+ | cosø-cos{ø+sin-1 (influence from the power factor when λ = 0 %/100)} | ]
± 1 digit when voltage and current are at the measurement range rated input
Accuracy of Phase Difference Φ ± [ | ø-cos-1(λ/1.0002) | + sin-1 (influence from the power factor when λ = 0 % / 100)] ± 1 digit when voltage and current are at the measurement range rated input
VOLTAGE, CURRENT AND ACTIVE POWER MEASUREMENTS
Item Specifications
Measurement Method Digital sampling method
Crest Factor 3 (voltage) or 4 (current)
Wiring System 1P2W, 1P3W, 3P3W, 3P4W, 3V3A
Peak Measurement Select manual or auto ranging
Auto Range Auto-range increase
The range is upped when any of the following conditions is met.
Vrms or Irms exceeds 130 % of the currently set measurement range.
Vpk, Ipk value of the input signal exceeds 300 % of the currently set measurement range.
Auto-range decline
The range is downed when all of the following conditions are met.
Vrms or Irms is less than or equal to 30 % of the measurement range.
Vrms or Irms is less than or equal to 125 % of the next lower measurement range.
Vpk, Ipk value of the input signal exceeds 300 % of the currently set measurement range.
Display Mode Switching AC+DC (the true RMS value of voltage and current)
V-MEAN (the rectified mean value calibrated to the RMS value of the voltage and the true RMS value of the
AC
DC
Measurement Synchronization Source Select voltage, current, or off
In the case of Auto Update Rate, select the voltage or current from the equipped element.
Line Filter Select OFF or ON (cutoff frequency at 5.5 kHz).
Peak Measurement Measures the peak (max, min) value of voltage, current or power from the instantaneous voltage, instantaneous current or instantaneous power that is sampled.
Zero-level Compensation Removes the internal offset of the measure unit (After measurement range is changed)
Measurement Parameters Voltage Vrms , Vmn, Vdc , Vac
Current Irms , Idc , Iac
Active Power P
Apparent Power VA
Reactive power VAR
Power Factor PF
Crest Factor CFI, CFV
Phase Angle DEG
Frequency IHz and VHz
Voltage Peak V+pk and V-pk
Current Peak I+pk and I-pk
Active Power Peak P+pk and P-pk
Total Harmonic Distortion THDI and THDV
Mathematical Computation MATH, EFFi
Maximum Current Ratio MCR
FREQUENCY MEASUREMENT
Item Specifications
Measurement Item Voltage and current
Measurement Frequency Range Data update interval Measurement Frequency Range
0.25 s 10 Hz ≤ f ≤ 10 kHz
0.5 s 5.0 Hz ≤ f ≤ 10 kHz
  1 s 2.0 Hz ≤ f ≤ 10 kHz
2 s 1.0 Hz ≤ f ≤ 10 kHz
5 s 0.5 Hz ≤ f ≤ 10 kHz
10 s 0.2 Hz ≤ f ≤ 10 kHz
20 s 0.1 Hz ≤ f ≤ 10 kHz
Auto ( * ) 0.1 Hz ≤ f ≤ 10 kHz
( * ) Limit of the measurement lower limit frequency by the Timeout setting
Timeout lower limit frequency
1 s 2.0 Hz
5 s 0.5 Hz
10 s 0.2 Hz
20 s 0.1 Hz
Measurement Range Auto switching among six types: 100 mHz, 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz.
Frequency Filter Select OFF or ON (cut off frequency of 5.5 kHz)
Accuracy Requirements When the input signal level is 30 % or more of the measurement range.
• Frequency filter is ON when measuring voltage or current of 200 Hz or less.
± (0.06 % of reading)
INTEGRATION
Item Specifications
Mode Select manual integration mode, standard integration mode, or repetitive integration mode.
Timer Automatically stop integration by setting a timer.
Selectable range: 0 hours 00 minutes 00 seconds to 9999 hours 59 minutes 59 seconds
Count Overflow WP: 999999 MWh / -99999 MWh
Accuracy ± (Power accuracy (or current accuracy) + 0.1 % of reading) (fixed range)
Range Setting Auto range or fixed range is available for Integration
Timer Accuracy ± 0.02 %
Remote Control Start, stop and reset operations are available using an external remote signal.
HARMONIC MEASUREMENT
Item Specifications
Measured Item Voltage, Current, Power
Measured Method Zero-cross simultaneous calculation method
Frequency Range 10 Hz to 1.2 kHz.
FFT Data Length 4096
(Auto switch when both (50 / 60) Hz and update rate must be greater than or equal to 0.25 s)
Sample Rate, Window Width,
and Upper Limit of Analysis Orders*
Fundamental Frequency Sample rate Window Width upper limit of Analysis orders
45 Hz to 55 Hz f × 512 10 50
54 Hz to 66 Hz f x 512 12 50
FFT data length 1024
Sample Rate, Window Width,
and Upper limit of Analysis Orders*
Fundamental Frequency Sample rate Window Width upper limit of Analysis orders
10 Hz to 67 Hz f × 1024 1 50
67 Hz to 150 Hz f × 512 2 32
150 Hz to 300 Hz f × 256 4 16
300 Hz to 600 Hz f × 128 8 8
600 Hz to 1200 Hz f × 64 16 4
Accuracy Frequency Voltage Current Power
10 Hz ≤ f < 45 Hz 0.15 % of reading 0.15 % of reading 0.35 % of reading
  + 0.35 % of range + 0.35 % of range + 0.50 % of range
  45 Hz ≤ f < 440 Hz 0.15 % of reading 0.15 % of reading 0.25 % of reading
  + 0.35 % of range + 0.35 % of range + 0.50 % of range
440 Hz ≤ f < 1.2 kHz 0.20 % of reading 0.20 % of reading 0.40 % of reading
  + 0.35 % of range + 0.35 % of range + 0.50 % of range
* 50 Hz / 60 Hz Compliant IEC61000-4-7 (update rate must be > 0.25 s).
* Harmonic calculation: FFT method in which FFT data length is divided into 2 types: 1024 and 4096.
* FFT data length automatically switches in accord with the Frequency and Update Rate of measured signal.
REMOTE CONTROL INPUT/OUTPUT SIGNAL
Item Specifications
Remote Control Input Signal Is_trig, Limit_trig
I/O Level TTL
I/O Logic Format Negative logic, Falling edge
Remote Control Out Ssignal M1_Pass, M2_Pass, M3_Pass, M4_Pass
I/O Level Relay form A
Output Rate Voltage 30 Vdc
I/O Sink Current Max. 0.5 A
* Q (VAR), S (VA), λ (PF) and Φ (DEG) are originated from the measured values including voltage, current and active power which go through computation process.
  In respect to distorted signal input, accordingly, the value acquired from other instruments, which employ different methods, may differ from that acquired from GPM-8340 unit.
* “Zero” will be shown for S or Q and “--“ will be displayed for λ and Φ when either current or voltage is less than 0.5 % of the rated range.

Specifications subject to change without notice.   PM-8340CD1BH


Do you have a Introduction of GPM-8340 4-Element AC Power Meter that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Introduction of GPM-8340 4-Element AC Power Meter calibrated today! Click Here!

 